Brookeld Engineering Labs., Inc. Page 77 Manual No. M13-2100
V.4.8 Calibration Reminder
TheDV3Tcanbeprogrammedtoprovidetheuserwithareminderthatcalibrationisdue.Two
parametersmustbeset:FrequencyinMonthsandStartDate.Whenbothparametersareentered,
theCalibrationReminderscreen(seeFigureV-13)willcalculateanddisplaythedaywhenthe
reminder will be presented.
ChecktheCalibrationReminderOncheckboxtoactivatethisfeature.
Figure V-13
TheCalibrationReminderwillbepresentedduringtherstpoweruponorafterthespeciedday.
Themessagebox(seeFigureV-14)requiresthattheuserpressOKbeforetheAutoZerowillbe
performed.
Figure IV-14
ThecalibrationreminderwillberesetbyBrookeldoryourauthorizedBrookelddealerwhenthe
unit is returned for calibration.