BROOKFIELD DV-I PRIMEDigital ViscometerOperating InstructionsManual No. M/07-022-D0613

Strona 2

Brookeld Engineering Laboratories, Inc. Page 10 Manual No. M/07-022-D0613I.4 SpecicationsSpeeds: 0.0, 0.3, 0.6, 1.5,

Strona 3 - TABLE OF CONTENTS

Brookeld Engineering Laboratories, Inc. Page 11 Manual No. M/07-022-D0613I.5 Setup1. To assemble the Model S Laboratory Stand, place the uprig

Strona 4 - (tear out page)

Brookeld Engineering Laboratories, Inc. Page 12 Manual No. M/07-022-D0613I.7 Key FunctionsFigure I-3 shows the control keys on the face of the

Strona 5

Brookeld Engineering Laboratories, Inc. Page 13 Manual No. M/07-022-D0613I.8 Preventative Maintenance and Cleaning Make sure the instrument i

Strona 6

Brookeld Engineering Laboratories, Inc. Page 14 Manual No. M/07-022-D0613 II. GETTING STARTEDII.1 Auto ZeroBefore readings may be taken, the

Strona 7

Brookeld Engineering Laboratories, Inc. Page 15 Manual No. M/07-022-D0613After approximately 15 seconds, the ashing stops and the following sc

Strona 8 - Figure I-1

Brookeld Engineering Laboratories, Inc. Page 16 Manual No. M/07-022-D0613S is blinking) the two character spindle value to the right of the S c

Strona 9 - Figure I-2

Brookeld Engineering Laboratories, Inc. Page 17 Manual No. M/07-022-D0613II.3 Speed Selection & Setting Table II-1 shows the available s

Strona 10

Brookeld Engineering Laboratories, Inc. Page 18 Manual No. M/07-022-D0613If the ARROW key is pressed just once and then released, the character

Strona 11 - PRIME must be earth

Brookeld Engineering Laboratories, Inc. Page 19 Manual No. M/07-022-D0613NOTE: If the motor is off or the RPM is 0.0, the maximum viscosity di

Strona 13

Brookeld Engineering Laboratories, Inc. Page 20 Manual No. M/07-022-D0613cP 78.0 S0110 RPM20 % 7.8Figure II-14Negative % (Torque) will be dis

Strona 14 - II. GETTING STARTED

Brookeld Engineering Laboratories, Inc. Page 21 Manual No. M/07-022-D0613II.9 Timed Modes for Viscosity Measurement The Timed Modes allow the

Strona 15

Brookeld Engineering Laboratories, Inc. Page 22 Manual No. M/07-022-D0613 Using the UP and DOWN ARROW keys, the user enters a value for the

Strona 16

Brookeld Engineering Laboratories, Inc. Page 23 Manual No. M/07-022-D0613%=76.4 RPM=100TIMED STOP DONEFigure II-22 The display will switch

Strona 17

Brookeld Engineering Laboratories, Inc. Page 24 Manual No. M/07-022-D0613TORQUE = 24.2%MIN: 15 SEC:13Figure II-25NOTE: When this mode has be

Strona 18 - RPM will blink for three

Brookeld Engineering Laboratories, Inc. Page 25 Manual No. M/07-022-D0613NOTE: For the Time to Torque method, the DV-I PRIME Viscometer will r

Strona 19

Brookeld Engineering Laboratories, Inc. Page 26 Manual No. M/07-022-D0613 When Temp Offset is enabled the temperature units display will be un

Strona 20

Brookeld Engineering Laboratories, Inc. Page 27 Manual No. M/07-022-D0613PRINT INTERVALSET MIN’S: 00Figure II-353. Use up and down arrows to

Strona 21

Brookeld Engineering Laboratories, Inc. Page 28 Manual No. M/07-022-D0613• Brookeld’s DV Loader software (for setting up test programs) is in

Strona 22

Brookeld Engineering Laboratories, Inc. Page 29 Manual No. M/07-022-D0613Figure II-39: Gather ScreenFigure II-40: DV Loader Screen

Strona 23

TABLE OF CONTENTS I. INTRODUCTION ...5I.1 Components

Strona 24 - UP or DOWN ARROW key

Brookeld Engineering Laboratories, Inc. Page 30 Manual No. M/07-022-D0613Figure II-41: Run/Data ScreenFigure II-42: Analysis Screen

Strona 25 -  TEMP OFFSET

Brookeld Engineering Laboratories, Inc. Page 31 Manual No. M/07-022-D0613II.12 Math ModelsMath models provide parameters that indicate how

Strona 26 - Cp 0.0 70.1°F

Brookeld Engineering Laboratories, Inc. Page 32 Manual No. M/07-022-D0613When n < 1 the product is shear-thinning or Pseudoplastic. This me

Strona 27

Brookeld Engineering Laboratories, Inc. Page 33 Manual No. M/07-022-D0613When n > 1 the product is shear-thickening or Dilatant. It’s appar

Strona 28

Brookeld Engineering Laboratories, Inc. Page 34 Manual No. M/07-022-D0613II.12.3 The Bingham Model€ t = € to+hD (€ t = shear stress, € to =

Strona 29 - Figure II-39: Gather Screen

Brookeld Engineering Laboratories, Inc. Page 35 Manual No. M/07-022-D0613II.12.4 The Casson Model€ t = to+ hD (€ t = shear stress, € to =

Strona 30 - Figure II-42: Analysis Screen

Brookeld Engineering Laboratories, Inc. Page 36 Manual No. M/07-022-D0613 An Example of the Casson Model at WorkBefore releasing a new over th

Strona 31

Brookeld Engineering Laboratories, Inc. Page 37 Manual No. M/07-022-D0613II.12.5 Other Common Rheological ModelsThe NCA/CMA Casson Model(1 +

Strona 32 - = shear stress

Brookeld Engineering Laboratories, Inc. Page 38 Manual No. M/07-022-D0613This model is a variation of the Power Law Model. Unlike the Power La

Strona 33

Brookeld Engineering Laboratories, Inc. Page 39 Manual No. M/07-022-D0613 III. MAKING VISCOSITY MEASUREMENTS III.1 Quick Start Viscosity Me

Strona 34

Brookeld Engineering Laboratories, Inc. Page 4 Manual No. M/07-022-D0613Appendix A - Cone/Plate Viscometer Set-Up ...

Strona 35

Brookeld Engineering Laboratories, Inc. Page 40 Manual No. M/07-022-D0613 When comparing data with others, be sure to specify the sample contai

Strona 36

Brookeld Engineering Laboratories, Inc. Page 41 Manual No. M/07-022-D0613increases. This behavior cannot be detected or evaluated with the sin

Strona 37

Brookeld Engineering Laboratories, Inc. Page 42 Manual No. M/07-022-D0613 Appendix A - Cone/Plate Viscometer Set-Up This Cone/Plate version of

Strona 38 - Figure II-49

Brookeld Engineering Laboratories, Inc. Page 43 Manual No. M/07-022-D0613A.2 Setup 1. Be sure that the Viscometer is securely mounted to the

Strona 39

Brookeld Engineering Laboratories, Inc. Page 44 Manual No. M/07-022-D0613A.3 Setting the Gap 1. Move the toggle switch to the right; this wil

Strona 40

Brookeld Engineering Laboratories, Inc. Page 45 Manual No. M/07-022-D0613A.4 Making Measurements with Cone/Plate Geometry Viscosity measuremen

Strona 41

Brookeld Engineering Laboratories, Inc. Page 46 Manual No. M/07-022-D0613A.5 Verifying Calibration1. Determine the appropriate sample volume.

Strona 42

Brookeld Engineering Laboratories, Inc. Page 47 Manual No. M/07-022-D0613 Appendix B - Viscosity Ranges LV (#1-4) and RV,HA,HB (#1-7) Viscomete

Strona 43 - CONE with the CUP!

Brookeld Engineering Laboratories, Inc. Page 48 Manual No. M/07-022-D0613Small Sample Adapter and ThermoselSSA and Thermosel SpindleViscosity

Strona 44 - Hit Point

Brookeld Engineering Laboratories, Inc. Page 49 Manual No. M/07-022-D0613DIN Adapter AccessoryDAA SpindleViscosity (cP)Shear Rate sec-1LVDV-I P

Strona 45

Brookeld Engineering Laboratories, Inc. Page 5 Manual No. M/07-022-D0613 I. INTRODUCTION The Brookeld DV-I PRIME Viscometer measures uid vis

Strona 46

Brookeld Engineering Laboratories, Inc. Page 50 Manual No. M/07-022-D0613When taking viscosity measurements with the DV-I PRIME Viscome

Strona 47 -   2)1cP=1mPa•s

Brookeld Engineering Laboratories, Inc. Page 51 Manual No. M/07-022-D0613 Appendix C - Variables in Viscosity Measurement As with any instrumen

Strona 48 - UL Adapter

Brookeld Engineering Laboratories, Inc. Page 52 Manual No. M/07-022-D0613Viscometer Related VariablesMost uid viscosities are found to be non-

Strona 49 - Helipath with T-Bar Spindles

Brookeld Engineering Laboratories, Inc. Page 53 Manual No. M/07-022-D0613 Appendix D - Spindle and Model CodesEach spindle has a two-digit c

Strona 50

Brookeld Engineering Laboratories, Inc. Page 54 Manual No. M/07-022-D0613Table D-2 lists the model codes and spring torque constants for each v

Strona 51

Brookeld Engineering Laboratories, Inc. Page 55 Manual No. M/07-022-D0613 Appendix E - Calibration Check Procedures For more help go to www.b

Strona 52

Brookeld Engineering Laboratories, Inc. Page 56 Manual No. M/07-022-D0613MINERAL OIL VISCOSITY STANDARD FLUIDS BEL Part No. Viscosity (cP) 2

Strona 53

Brookeld Engineering Laboratories, Inc. Page 57 Manual No. M/07-022-D0613 Follow theses steps using one of the recommended spindles to verify c

Strona 54

Brookeld Engineering Laboratories, Inc. Page 58 Manual No. M/07-022-D0613Calibration Check Procedure for a Thermosel SystemA two-step process i

Strona 55 - +/-) 0.1°C

Brookeld Engineering Laboratories, Inc. Page 59 Manual No. M/07-022-D0613Calibration Check Procedures for DIN Adapter When a DIN UL Adapter is

Strona 56

Brookeld Engineering Laboratories, Inc. Page 6 Manual No. M/07-022-D0613The equivalent units of measurement in the SI system are calc

Strona 57

Brookeld Engineering Laboratories, Inc. Page 60 Manual No. M/07-022-D0613Calibration Check Procedure for Cone/Plate Viscometers1) Follow the p

Strona 58

Brookeld Engineering Laboratories, Inc. Page 61 Manual No. M/07-022-D0613Interpretation of Calibration Test Results:When verifying the calibrat

Strona 59

Brookeld Engineering Laboratories, Inc. Page 62 Manual No. M/07-022-D0613 Appendix F - The Brookeld GuardlegThe guard leg was originally des

Strona 60

Brookeld Engineering Laboratories, Inc. Page 63 Manual No. M/07-022-D0613repeatable value can be obtained without any special effort for any me

Strona 61

Brookeld Engineering Laboratories, Inc. Page 64 Manual No. M/07-022-D0613 Appendix G - CommunicationsDV-I PRIMESerial and Analog Outputs-2345

Strona 62

Brookeld Engineering Laboratories, Inc. Page 65 Manual No. M/07-022-D0613 Appendix H - Laboratory Stands with Parts IdenticationModel S

Strona 63

Brookeld Engineering Laboratories, Inc. Page 66 Manual No. M/07-022-D0613UNPACKINGCheck carefully to see that all the components are received w

Strona 64

Brookeld Engineering Laboratories, Inc. Page 67 Manual No. M/07-022-D0613 Appendix I - DVE-50A Probe ClipProbe Clip DVE-50A is supplied with t

Strona 65 - Release Lever: Push

Brookeld Engineering Laboratories, Inc. Page 68 Manual No. M/07-022-D0613 Appendix J - Fault Diagnosis and Troubleshooting Spindle Does Not Ro

Strona 66

Brookeld Engineering Laboratories, Inc. Page 69 Manual No. M/07-022-D0613 Appendix K - Online Help and Additional Resourceswww.brookeldengine

Strona 67 - Temperature Probe

Brookeld Engineering Laboratories, Inc. Page 7 Manual No. M/07-022-D0613I.2 UtilitiesInput Voltage: 115 VAC or 230 VACInput Frequency: 50/60

Strona 68

Brookeld Engineering Laboratories, Inc. Page 70 Manual No. M/07-022-D0613 Appendix L- Warranty Repair and ServiceWarrantyBrookeld Viscometer

Strona 69

Brookeld Engineering Laboratories, Inc. Page 71 Manual No. M/07-022-D0613 Appendix M - Viscosity Test Report

Strona 70

Brookeld Engineering Laboratories, Inc. Page 8 Manual No. M/07-022-D0613I.3 Components and DimensionsCone/Plate OptionTemperature Probe Option

Strona 71

Brookeld Engineering Laboratories, Inc. Page 9 Manual No. M/07-022-D0613Figure I-2