Brookeld Engineering Labs., Inc. Page 68 Manual No. M13-2100
Figure V-8
The RunningViscosityTestdisplay will show acolored boxaround thedata
parameterassociatedwiththeGlobalAlarm.Whenthealarmconditionisrst
met,thecoloredboxwillbeyellow.Ifthemeasurementparameterreturnstothe
acceptablerange,thenthecoloredboxwillberemoved.Whenthealarmcondition
ismetasecondtime,thecoloredboxwillbered(seeFigureV-9).Theaudible
alarm will only sound once.
Figure V-9
TIP: TheGlobalAlarmsoundcanbeturnedoffintheSoundsmenuofUser
Settings.
The GlobalAlarm is deactivated by choosing None in the drop down list of
availablemeasurementparameters.YoumustpressOKtoconrmyourGlobal
Alarmchoice.