BROOKFIELD RS-CPS+ RHEOMETER
Operating Instructions
Manual No. M08-218-B0212
(for serial numbers beginning with “304”)
SPECIALISTS IN THE
MEASUREMENT AND
CONTROL OF VISCOSITY
T
EL
508-946-6200
F
AX
508-946-6262
or 800-628-8139 (USA excluding MA)
I
NTERNET
http://www.brookfieldengineering.com
BROOKFIELD ENGINEERING LABORATORIES, INC.
11 Commerce Boulevard, Middleboro, MA 02346 USA
with offices in
:
Boston Chicago London Stuttgart Guangzhou
Przeglądanie stron 0
1 2 3 4 5 6 ... 61 62

Podsumowanie treści

Strona 1 - Manual No. M08-218-B0212

BROOKFIELD RS-CPS+ RHEOMETEROperating InstructionsManual No. M08-218-B0212(for serial numbers beginning with “304”) SPECIALISTS IN THEMEASUREMEN

Strona 2

BrookeldEngineeringLabs.,Inc. Page10 ManualNo.M08-218-B0212III.2.1 TemperatureSensorThe connecting cable of the temperature sensor Pt 100

Strona 3 - Table of Contents

BrookeldEngineeringLabs.,Inc. Page11 ManualNo.M08-218-B0212III.2.4 ComputerConnectionIf the R/S-CPS+ Rheometer has to be used in “REMOTE” m

Strona 4

BrookeldEngineeringLabs.,Inc. Page12 ManualNo.M08-218-B0212The following working uids are usually used in the temperature bath:-10°C to +90°

Strona 5 - I.2 MeasuringPrinciple

BrookeldEngineeringLabs.,Inc. Page13 ManualNo.M08-218-B0212Peliter Control Cable INPeltier Power Supply INDrawer G Fuse5x20 1A (delay) Main S

Strona 6 - II.1 R/S-CPS+ Rheometer

BrookeldEngineeringLabs.,Inc. Page14 ManualNo.M08-218-B0212 IV.Environment,Handling,CleaningandMaintenance IV.1 OperatingEnvironmen

Strona 7 - R/S-CPS+Rheometer

BrookeldEngineeringLabs.,Inc. Page15 ManualNo.M08-218-B0212 IV.4 MaintenanceThe R/S Rheometer system is designed for long-term operation. S

Strona 8 - II.3 ComputerSystem

BrookeldEngineeringLabs.,Inc. Page16 ManualNo.M08-218-B0212 V.MeasuringSystemsThis section covers the various measurement geometries. Th

Strona 9 - III.2 Electrical Connections

BrookeldEngineeringLabs.,Inc. Page17 ManualNo.M08-218-B0212Fig. 6: Setting the gap on the R/S-CPS+ RheometerSet the micrometer ring (Nonius

Strona 10 - III.2.3 Printer Connection

BrookeldEngineeringLabs.,Inc. Page18 ManualNo.M08-218-B0212 V.2 FillingtheMeasuringSystemsPlace the sample material on the bottom plate

Strona 11 - III.2.4 ComputerConnection

BrookeldEngineeringLabs.,Inc. Page19 ManualNo.M08-218-B0212 VI.OperationandMenuSystemThe following section summarizes the operation and

Strona 13

BrookeldEngineeringLabs.,Inc. Page20 ManualNo.M08-218-B0212 VI.1 KeyboardSTOKR/S + RHEOMETERONMAIN MENU - Single measurement - Program measu

Strona 14 -  IV.2 Handling

BrookeldEngineeringLabs.,Inc. Page21 ManualNo.M08-218-B0212Key Operation FunctionofKey ExampleST1) Menu2) Input of numerical values3) Sele

Strona 15 - IV.4 Maintenance

BrookeldEngineeringLabs.,Inc. Page22 ManualNo.M08-218-B0212 VI.2 MenuSystemofR/S-CPS+Rheometer MainMenuSingleMeasurementProgramMeasurem

Strona 16

BrookeldEngineeringLabs.,Inc. Page23 ManualNo.M08-218-B0212Menu HandlingSince the LCD of the R/S CPS+ Rheometer cannot show all menu items si

Strona 17 - MicrometerRing

BrookeldEngineeringLabs.,Inc. Page24 ManualNo.M08-218-B0212Example:Select Meas.system:1) C25-2/302) C50-1/303) C50-2/30 VI.4 InputofNumeri

Strona 18

BrookeldEngineeringLabs.,Inc. Page25 ManualNo.M08-218-B0212Input of Values:Val.[1/s]: 0290.00Nr. of MP: 010Time[s]: 0100 Note: If you press

Strona 19

BrookeldEngineeringLabs.,Inc. Page26 ManualNo.M08-218-B0212THERMOSTATTS: 20.0T: 20.0Status: 00 Permanent Indication ofActual Temperature Val

Strona 20 - VI.1 Keyboard

BrookeldEngineeringLabs.,Inc. Page27 ManualNo.M08-218-B0212Values and Alphanumeric Texts)• Input of the measurement duration in seconds. Th

Strona 21

BrookeldEngineeringLabs.,Inc. Page28 ManualNo.M08-218-B0212This message will be displayed until the rst measuring point is reached and displ

Strona 22

BrookeldEngineeringLabs.,Inc. Page29 ManualNo.M08-218-B0212The Run Program function starts one of the available programs dened in the memory

Strona 23 -  VI.3 SelectingfromLists

Table of Contents I. General Description ...5 I.1 Use

Strona 24

BrookeldEngineeringLabs.,Inc. Page30 ManualNo.M08-218-B0212Measurements can be taken with the memory full but the results will only be shown

Strona 25

BrookeldEngineeringLabs.,Inc. Page31 ManualNo.M08-218-B0212VI.5.4 MAINMenu→ RemoteThe “Remote” function initiates measurements to be made

Strona 26

BrookeldEngineeringLabs.,Inc. Page32 ManualNo.M08-218-B0212 VI.6 MenuEntriesintheUtilitiesMenuThe Utilities menu contains several usef

Strona 27

BrookeldEngineeringLabs.,Inc. Page33 ManualNo.M08-218-B0212VI.6.2 Utilities → EditProgramThis function allows the input of new programs and

Strona 28

BrookeldEngineeringLabs.,Inc. Page34 ManualNo.M08-218-B0212The rst user dened value ( = the rst measuring point) is always the start value

Strona 29 - Automatic increment at

BrookeldEngineeringLabs.,Inc. Page35 ManualNo.M08-218-B0212AvailableMeasurementTypes UserDenedRangeShear Rate D{s-1] Depends on the meas

Strona 30

BrookeldEngineeringLabs.,Inc. Page36 ManualNo.M08-218-B0212Maximum step duration: 3600 sThe instrument will automatically check the input.

Strona 31 - VI.5.5 MAINMenu→ Utilities

BrookeldEngineeringLabs.,Inc. Page37 ManualNo.M08-218-B0212ERROR #1Printer not ready!<OK>cont. <ST>stopIf this error message app

Strona 32

BrookeldEngineeringLabs.,Inc. Page38 ManualNo.M08-218-B0212<OK>=storingName: C50-1/30tau_prom: 01.1418k_gamma: 01.2910Use the OK ke

Strona 33 - D[1/s]=f(t)

BrookeldEngineeringLabs.,Inc. Page39 ManualNo.M08-218-B0212Before deleting, ensure that the data has been archived to a printer or PC. Befor

Strona 34

VI.6.1 Utilities → Zero Calibration ...32VI.6.2 Utilities → Edit Program ...

Strona 35

BrookeldEngineeringLabs.,Inc. Page40 ManualNo.M08-218-B0212VI.7.1 Conguration→ OutputModeThis function denes the output device (printer,

Strona 36 - <OK>=Storing

BrookeldEngineeringLabs.,Inc. Page41 ManualNo.M08-218-B0212Select meascount mode:-No resetting-Reset every dayResetting only via func-tion me

Strona 37

BrookeldEngineeringLabs.,Inc. Page42 ManualNo.M08-218-B0212VI.7.5 Conguration→RS232 ParametersThis function allows the preselection of int

Strona 38 - Time of Measurement

BrookeldEngineeringLabs.,Inc. Page43 ManualNo.M08-218-B0212When all parameters have been selected, the new parameters are displayed:RS232: 1

Strona 39

BrookeldEngineeringLabs.,Inc. Page44 ManualNo.M08-218-B02121. Switch the rheometer and the PC off.2. Connect the rheometer (RS232 connector

Strona 40

BrookeldEngineeringLabs.,Inc. Page45 ManualNo.M08-218-B0212 VII. Measuring VII.1 MeasuringinManualOperationYou can measure in manual mo

Strona 41

BrookeldEngineeringLabs.,Inc. Page46 ManualNo.M08-218-B0212 VIII. Technical DataR/S-CPS+ RheometerDimensions 480 mm x 300 mm x 290 mmWeigh

Strona 42 - Databits [Bit] 7 or 8

BrookeldEngineeringLabs.,Inc. Page47 ManualNo.M08-218-B0212Value Symbol Physical UnitSpeed n [min-1]Torque (relative)(1000 ‰ = 50 mNm) M [1}T

Strona 43 - <OK>=storing

BrookeldEngineeringLabs.,Inc. Page48 ManualNo.M08-218-B0212 Appendix A A.1 DataSheetsofStandardMeasuringSystemsTable A-1: Cone/Plate

Strona 44

BrookeldEngineeringLabs.,Inc. Page49 ManualNo.M08-218-B0212 A.2 ErrorMessagesThe R/S-CPS+ Rheometer is user-friendly regarding errors. Wh

Strona 45

BrookeldEngineeringLabs.,Inc. Page5 ManualNo.M08-218-B0212 I. General DescriptionThis section has general information about the instrument

Strona 46 - VIII. Technical Data

BrookeldEngineeringLabs.,Inc. Page50 ManualNo.M08-218-B0212 Cause: An impermissibly high value was measured during the zero point cali-bratio

Strona 47 - (1000 ‰ = 50 mNm)

BrookeldEngineeringLabs.,Inc. Page51 ManualNo.M08-218-B0212Cause: The maximum speed has been exceeded during a measurement. Possible reasons

Strona 48

BrookeldEngineeringLabs.,Inc. Page52 ManualNo.M08-218-B0212The AC adapter unit of the R/S-CPS+ Rheometer permits operation of the rheometer w

Strona 49

BrookeldEngineeringLabs.,Inc. Page53 ManualNo.M08-218-B0212Othercountries• The connectors of the mains connecting cable must be approved a

Strona 50

BrookeldEngineeringLabs.,Inc. Page54 ManualNo.M08-218-B0212• R/S-CPS+ Cone/Plate Rheometer with appropriate cables• Temperature control app

Strona 51 - Rheometer

BrookeldEngineeringLabs.,Inc. Page55 ManualNo.M08-218-B02121. Turn on the R/S-CPS+ Rheometer.2. Turn on the temperature controller. Set th

Strona 52

BrookeldEngineeringLabs.,Inc. Page56 ManualNo.M08-218-B021211. Manually turn the outside ring of the dial indicator so that the needle is on

Strona 53

BrookeldEngineeringLabs.,Inc. Page57 ManualNo.M08-218-B0212The R/S-CPS+ calibration check can be performed with or without software. If soft

Strona 54

BrookeldEngineeringLabs.,Inc. Page58 ManualNo.M08-218-B0212 StepNr1Start [‰] :250End [‰] :250Nr. Of Mp: 40time [s] : 60StepNr2Start

Strona 55 -  B.3 SettingtheGap

BrookeldEngineeringLabs.,Inc. Page59 ManualNo.M08-218-B0212Other FaultsIndication PossibleCause SolutionDial indicator reading is inconsiste

Strona 56

BrookeldEngineeringLabs.,Inc. Page6 ManualNo.M08-218-B0212 II. SystemCongurationThe R/S CPS+ Rheometer system consists of:• Rheometer he

Strona 57

BrookeldEngineeringLabs.,Inc. Page60 ManualNo.M08-218-B0212B.5.5 Cone Spindle Data Example

Strona 58

BrookeldEngineeringLabs.,Inc. Page61 ManualNo.M08-218-B0212www.brookeldengineering.comThe Brookeld website is a good resource for additiona

Strona 59

BrookeldEngineeringLabs.,Inc. Page62 ManualNo.M08-218-B0212Brookeld Viscometers are guaranteed for one year from date of purchase against de

Strona 60

BrookeldEngineeringLabs.,Inc. Page7 ManualNo.M08-218-B0212R/S-CPS+Rheometer1 Measuring Head of R/S-CPS+ Rheometer2 Measuring Element Coupl

Strona 61

BrookeldEngineeringLabs.,Inc. Page8 ManualNo.M08-218-B0212 II.2 MeasuringDevicesThe basic instrument is equipped with a temperature contro

Strona 62

BrookeldEngineeringLabs.,Inc. Page9 ManualNo.M08-218-B0212 III.InstrumentInstallationIn this section you will learn how to set up and use t

Komentarze do niniejszej Instrukcji