Brookeld Engineering Labs., Inc. Page 67 Manual No. M13-2100
Regional Settings:TheRegionalSettingsmenudisplaysthecurrentselectionsforthedefault
language,numberformat,dateformatanddayseparator.Eachsettingcan
bechangedbypressingthebluedownarrowandselectingthedesiredvalue.
RegionalSettingsarepartofthedataleforsavedresults.
TheLanguagesettingestablishedthedefaultlanguagefortheDV3T.Individual
userscanchooseadifferentlanguageforusewiththeirspeciclogin(see
SectionV.2).
TheformatsforNumber,Time,DaySeparatorandDateofferseveraloptions.
Number Format 9,999,999.00 USA decimal convention
9.999.999,00 Europe comma convention
9999999,00 SI/ISO space convention
TimeFormat 13:59:59 24 Hour Format
1:59:59 PM 12 Hour Format
Day Separator / ForwardSlash
- Dash
. Period
Date Format Year/Month/Day ISO Standard
Month/Day/Year USA Standard
Day/Month/Year International Standard
Global Alarm: TheGlobalAlarmestablishesasinglemeasurementparameterwitharangeof
acceptance.Whenever the measured value exceedsthis rangeof acceptance,
anaudibleandvisualalarmwillbeissued.TheGlobalAlarmwillapplytoany
testthatisrun.TheQCLimitalsoestablishesameasurementparameterwitha
rangeofacceptance;however,theQCLimitisappliedonlytothetestinwhich
itisdened.
TheGlobalAlarmcanbesetforanyofthefourmeasurementparametersincluding;
ShearStress,Temperature,TorqueandViscosity.Onceaparameterisselected,
bothlowandhighlimitsmustbeentered.Theselimitsareindependentofthe
spindleandspeedselected.Whentheglobalalarmistriggered,theDV3Twill
provideanaudiblealarmandaWarningmessageonthescreen(seeFigureV-8).
Theerrormessagewillprovidesomedetailaboutthealarmcondition.Thetest
anddatacollectionwillcontinuewhiletheWarningmessageisdisplayed.The
audiblealarmwillcontinuetosounduntiltheoperatortouchestheOKbutton