Brookfield DV3T Rheometer Instrukcja Użytkownika Strona 112

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 125
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 111
Brookeld Engineering Labs., Inc. Page 112 Manual No. M13-2100
3) Totalaccuracyisthesumofthevaluesn(1)and(2):
At6RPM,accuracyis:15.6cP+1.0cP=+/-16.6cP
At12RPM,accuracyis:7.8cP+1.0cP=+/-9.8cP
At30RPM,accuracyis:3.1cP=1.0cP=+/-4.1cP
4) Therefore,ateachspeed,theacceptablewindowswithinwhichthemeasuredviscosityvalue
mustlieiscalculatedrelativetotheviscosityvalueofthestandard:
At6RPM:84.9cPto118.1cP
At12RPM:91.7cPto111.3cP
At30RPM:97.4cPto105.6cP
Ifyourmeasuredvaluesfalloutsideofthesewindows,contactBrookeldoryourauthorized
dealertodiscussyourresultsanddeterminewhetheryourinstrumentisoutofcalibration.
Przeglądanie stron 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 124 125

Komentarze do niniejszej Instrukcji

Brak uwag