Brookeld Engineering Labs., Inc. Page 120 Manual No. M13-2100
Appendix J - Fault Diagnosis and Troubleshooting
ListedaresomeofthemorecommonproblemsthatyoumayencounterwhileusingyourRheometer.
❏ Spindle Does Not Rotate
✓Makesuretherheometerispluggedin.
✓Checkthevoltageratingonyourrheometer(115,220V);itmustmatchthewallvoltage.
✓MakesurethemotorisONandthedesiredrpmisselected.
❏ Spindle Wobbles When Rotating or Looks Bent
✓Makesurethespindleistightenedsecurelytotherheometercoupling.
✓Checkthestraightnessofallotherspindles;replaceifbent.
✓Inspectrheometercouplingandspindlecouplingmatingareasandthreadsfordirt;cleanthreads
onspindlecouplingwitha3/56left-handtap.
✓Inspectthreadsforwear;ifthethreadsareworn,theunitneedsservice(seeAppendixM).Check
toseeifspindlesrotateeccentricallyorwobble.Thereisanallowablerunoutof1/32-inchin
eachdirection(1/16-inchtotal)whenmeasuredfromthebottomofthespindlerotatinginair.
✓Checktoseeiftherheometercouplingappearsbent;ifso,theunitisinneedofservice(see
AppendixM:WarrantyRepairandService).
❏ Inaccurate Readings
✓Verifyspindle,speedandmodelselection.
✓VerifyspindleselectioniscorrectonDV3T.
✓ If%readingsareunder-range(lessthan10%),thedisplaywillash;changespindle and/or
speed.
✓“EEEE”onthedigitaldisplaymeanstheunitisover-range(greaterthan100%);reducespeed
and/orchangespindle.
✓ Verifytestparameters:temperature,container,volume,method.Referto:
• “MoreSolutionstoStickyProblems”,SectionIII
• “DV3TRheometerOperatingInstructions”,AppendixC:VariablesinViscosityMeasurements
✓Performacalibrationcheck;followtheinstructionsinAppendixE.
✓ Verifytolerancesarecalculatedcorrectly.
✓Verifythecalibrationcheckprocedureswerefollowedexactly.
Iftheunitisfoundtobeoutoftolerance,theunitmaybeinneedofservice.SeeAppendixM
fordetailsonWarrantyRepairandService.