Brookeld Engineering Labs., Inc. Page 103 Manual No. M13-2100
Appendix D - Spindle and Model Codes
EachspindlehasatwodigitentrycodewhichisenteredviathekeypadontheDV3T.Theentry
codeallowstheDV3TtocalculateViscosity,ShearRateandShearStressvalues.
Eachspindlehastwoconstantswhichareusedinthesecalculations.TheSpindleMultiplier
Constant(SMC)usedforviscosityandshearstresscalculations,andtheShearRateConstant
(SRC),usedforshearrateandshearstresscalculations.NotethatwhereSRC=0,noshearrate/
shearstresscalculationsaredoneandthedatadisplayediszero(0)forthesefunctions.
Table D-1
(Continued on following page)
SPINDLE ENTRY CODE SMC SRC
RV1 01 1 0
RV2 02 4 0
RV3 03 10 0
RV4 04 20 0
RV5 05 40 0
RV6 06 100 0
RV7 07 400 0
HA1 01 1 0
HA2 02 4 0
HA3 03 10 0
HA4 04 20 0
HA5 05 40 0
HA6 06 100 0
HA7 07 400 0
HB1 01 1 0
HB2 02 4 0
HB3 03 10 0
HB4 04 20 0
HB5 05 40 0
HB6 06 100 0
HB7 07 400 0
LV1 61 6.4 0
LV2 62 32 0
LV3 63 128 0
LV4 or 4B2 64 640 0
LV5 65 1280 0
LV-2C 66 32 0.212
LV-3C 67 128 0.210
Spiral 70 105 0.677