Brookfield DV3T Rheometer Instrukcja Użytkownika Strona 121

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 125
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 120
Brookeld Engineering Labs., Inc. Page 121 Manual No. M13-2100
Rheometer Will Not Return to Zero
Rheometerisnotlevel
• Checkwithspindleoutofthesample.
• Adjustthelaboratorystand.
Pivotpointorjewelbearingfaulty
• PerformanOscillationCheck*
RemovethespindleandturnthemotorOFF;selectdisplayto%torquemode.
Gentlypushupontherheometercoupling.
Turnthecouplinguntilthedigitaldisplayreads10-15onthe%display.
Gentlyletgoofthecoupling.
Watchthedigitaldisplay;youshouldseea“run”ofnumbersnexttothe%;the“run”of
numbersshouldultimatelystopat0.0(+/-0.1).
IfthedigitaldisplaydoesnotreturntoZERO,theunitmostlikelyisinneedofservice.
• Performcalibrationcheck(seeAppendixE).
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerforrepair(see
AppendixM).
*Thisproceduredoesnotapplytoinstrumentswithballbearingsuspension(seeSectionI.3).
Display Reading Will Not Stabilize
Specialcharacteristicofsampleuid.Thereisnoproblemwiththerheometer.
• RefertoAppendixC.
Checkforerraticspindlerotation.
• Verifypowersupply
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerforrepair.
Bent spindle or spindle coupling.
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerforrepair.
Temperatureuctuationinsampleuid.
• Usetemperaturebathforcontrol.
Przeglądanie stron 120
1 2 ... 116 117 118 119 120 121 122 123 124 125

Komentarze do niniejszej Instrukcji

Brak uwag