
Brookeld Engineering Labs., Inc. Page 20 Manual No. M13-2100
ThecommandKeysincludeClear,SaveandRun.
Clear: Clearalldatathathasbeenenteredintothetestparameters
andrestorethevaluestothefactorydefault.
Save: Savethecurrenttest.
Run: Runthecurrenttest.
Thetestparameterareaincludesmanyelementsoftheyieldtestaswellaslivemeasurementof
temperature.TemperaturedatawillonlybedisplayedifaBrookeldtemperatureprobeisconnected
totheDV3TRheometer.
Spindle: Thecurrentlyselectedspindle.AllYieldandStresscalculations
willbemadebasedontheSpindleselectionandtheImmersion
Mark.Spindlechoiceislimitedtovanespindlesoruserdened
custom spindles.
Immersion Mark: SpecifytheimmersionlocationoftheBrookeldvanespindle.
Primaryisattheimmersionmarkshownonthespindleshaft.
Secondaryisatthemid-pointofthevanesurface(seeSection
IV.5).
Pre-ShearSpeed: Specifyaspedofrotationforpre-shearofthesample
Pre-ShearTime: SpecifythedurationofthePre-Shearstep.
Zero Speed: Specifythespeedofrotationusedtoreturnthe%Torquereading
tozerobeforestartingtheyieldmeasurement.
WaitTime: Specifythedurationofarestperiodpriortostartingtheyieldtest.
RunSpeed: Specifythespeedoftheyieldtest.
TorqueReduction: Specifytheendpointofthetest.Normallysetto100%.
Temperature: Alivesignalfromtherheometerwhenatemperatureprobe
is attached (Brookeld Part No. DVP-94Y). Also shown
is the set point used to control the attached temperature
controller(BrookeldThermoselorTC-XX0SD/TC-XX0AP
TemperatureBaths).
YieldLimit: Denethelimitsforacceptablemeasurementdata.
Instructions: Createamessagethattheuserwillseewhenthetestbegins.
SavePath: Denethememorylocationwherethedatawillbesaved.
Komentarze do niniejszej Instrukcji