Brookeld Engineering Labs., Inc. Page 77 Manual No. M13-167-B0614
However,instrumentaccuracyis±2%ofthemaximumviscosityrangeandnot
thestandard1%.
Calibration Procedure for Cone/Plate Viscometers
1) Followtheaboveproceduresformechanicallyadjustingthesettingoftheconespindle
totheplate.
2) RefertoAppendixA;TableA-1,anddeterminethecorrectsamplevolumerequired
fortheselectedspindle.
3) Selectaviscositystandarduidthatwillgiveviscosityreadingsbetween10%and
100%offullscalerange.RefertoAppendixBforviscosityrangesofconespindles.
ConsultwithBrookeldoranauthorizeddealertodeterminewhichuidisappropriate.
Itisbesttouseaviscositystandarduidthatwillbeclosetothemaximumviscosity
for a given cone spindle/speed combination.
Example:
DV2TLVViscometer,ConeCP-42,Fluid10
 Havingaviscosityof9.7cPat25°C
At60RPM,thefullscaleviscosityrangeis10.0cP.Thus,theViscometerreading
shouldbe97%torqueand9.7cPviscosity±0.197cP(0.1cPfortheviscometerplus
0.097cPfortheuid).TheaccuracyisacombinationofViscometeranduidtolerance
(refertoInterpretationofCalibrationTestResults).
4) Withtheviscometerstopped,removethesamplecupandplacetheviscositystandard
uidintothecup,waiting10minutesfortemperatureequilibrium.
5) ConnectthesamplecuptotheViscometer.Allowsufcienttimefortemperatureto
reachequilibrium.Typically15minutesisthemaximumtimethatyoumustwait.Less
timeisrequiredifspindleandcuparealreadyattesttemperature.
6) MeasuretheviscosityandrecordtheViscometerreadinginboth%torqueandcentipoise
(cP).
Notes: 1) The spindle must rotate at least ve (5) times before a viscosity
reading is taken.
2) TheuseofBrookeldViscosityStandarduidsintherangeof5cPto
5000 cP is recommended for cone/plate instruments. Please contact
Brookeld Engineering Laboratories or an authorized dealer if your
calibrationprocedurerequiresmoreviscousstandards.
3) Selectaviscositystandarduidthatwillgiveviscosityreadingsbetween
10%and100%of fullscalerange.Refer toAppendixBforviscosity
rangesofconespindles.Donotuseasiliconeviscositystandarduid
withaviscosityvaluegreaterthan5000cPwithaCone/PlateViscometer.
Brookeld offers a complete range of mineral oil viscosity standards
suitable for use with Cone/Plate Viscometersas shown inTable E-2.
ConsultwithBrookeldoranauthorizeddealertodeterminewhichuid
is appropriate.
Interpretation of Calibration Test Results:
WhenverifyingthecalibrationoftheDV2T,theinstrumentandviscositystandarduid
errormustbecombinedtocalculatethetotalallowableerror.
TheDV2Tisaccurateto(+/-)1%oftherangeinusewhenusingspindlesLV#1-3,RV
#2-6,HA#2-6andHB#2-6.WhenusinganaccessorydevicewiththeDV2Tsuchas