Brookfield DV2T Viscometer Instrukcja Użytkownika Strona 53

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 95
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 52
Brookeld Engineering Labs., Inc. Page 53 Manual No. M13-167-B0614
IV.4.5 Default Path
ThePathisthelocationwheredataortestsarestored.TheDV2Twillallowforstorage
totheInternalMemoryortoanattachedUSBFlashdrive.Thepathwillalsoincludeany
lestructurethathasbeencreated.
TheDefaultPathidentieswhatlocationwillbeusedastheinitiallocationwhenResultsare
saved. The Default Path dened in Reports - Congure ViscosityTest (saved test or
unsavedtest)willtakepriorityovertheDefaultPath.
TheDV2TcancreatelestructurebasedonUserIDandTestName.Theoptionsprovided
fortheDefaultPathspecifywhatformatthedatapathshouldutilize.Forexample:set
theDefaultPathtoOperator->TestName.TheUserisloggedinasDAVIDandthetest
torunisINK5.ThedatapathwillbedefaultedasshowninFigureIV-10.
Figure IV-10
TIP:TheDefaultPathcanbeoverwrittenatthetimethatdataissaved.
IV.4.6 Settings Reset
SettingsResetwillreturntheDV2Tsettings(Device,UserandGlobal)tothefactorydefault.
Internal memory will not be affected. User accounts will not be affected.
IV.4.7 Device Reset
DATA LOSS POSSIBLE!
DeviceResetwillreturntheDV2Ttothefactorydefaultcondition.Alldata,testsandAudit
Trailswillbedeletedfrominternalmemory.AllUseraccountswillbedeleted.
TIP:MovedatalesandtestlestoaUSBUSBFlashDrivepriortoselectingDevice
Reset.
Przeglądanie stron 52
1 2 ... 48 49 50 51 52 53 54 55 56 57 58 ... 94 95

Komentarze do niniejszej Instrukcji

Sachin Vaijanath Ashture 05 Mar 2025 | 15:48:33

It's good information