Brookfield DV2+Pro Viscometer Instrukcja Użytkownika Strona 87

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 101
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 86
Brookeld Engineering Labs., Inc. Page 87 Manual No. M03-165-F0612
should be known that this type of torque reading will not convert into a correct centipoise value when
usingaBrookeldfactoriftheboundaryconditionsarenotthosespeciedbyBrookeld.
TheguardlegisapartofthecalibrationcheckoftheBrookeldLVandRVseriesViscometer/Rheometer.
Our customers should be aware of its existence, its purpose and the effect that it may have on data.
Withthisknowledge,theviscometerusermaymakemodicationstotherecommendedmethodof
operation to suit their needs.
4 11/16
1 7/163 3/16
5 9/32
Figure F-1 - Brookeld Guard Leg
B-21KY
Guard Leg
B-20KY
Guard Leg
5 7/8
5 5/32
For RV Torque
For LV Torque
3 3/16
1 7/16
Figure F-2 - Brookeld EZ-Lock Guard Leg