Brookeld Engineering Labs., Inc. Page 8 Manual No. M03-165-F0612
I.2 Utilities
InputVoltage: 115VACor230VAC
InputFrequency: 50/60Hz
PowerConsumption: 30VA
PowerCordColorCode:
United States Outside United States
Hot (live) Black Brown
Neutral White Blue
Ground(earth) Green Green/Yellow
Mainsupplyvoltageuctuationsarenottoexceed±10%ofthenominalsupply
voltage.
I.3 Specications
Speeds: Choice of 3 options.
Instrument has “Interleaved” speeds when manufactured.
Standalone Interleaved: LV/RV(18speeds)8LVspeedsfollowedby
10 RV speeds
Sequential: LV/RV (18 speeds) 8 LV speeds and 10 RV
speeds arranged in sequential order from lowest
0.3 rpm to highest 100 rpm.
Custom: 54speeds,userselectable
External (PC Control) 0.01 - 200 rpm
0.01 rpm increments from 0.01 to 0.99 rpm
0.1 rpm increments from 1.0 to 200 rpm
Note: Refer to Appendix G for detailed list of all speeds.
Weight: Gross Weight 23 lbs. 10.5 kg.
Net Weight 20 lbs. 9 kg.
Carton Volume 1.65 cu. ft. 0.05 m
3
Carton Dimensions 22 in. (56 cm) W x 11 in. (28 cm) L x 22 in. (56 cm) H
Temperature Sensing Range: -100˚Cto300˚C(-148˚Fto572˚F)
Analog Torque Output: 0 - 1 Volt DC (0 - 100% Torque)
Analog Temperature Output: 0 - 3.75 Volts DC (-100°C to +275°C)
RS232 Compatible Serial Port for use with an attached printer or PC.
Centronics Compatible Parallel Port for use with an attached printer.
Viscosity Accuracy:
±1.0%offullscalerange
The use of accessory items will have an effect on the
measurement accuracy. See Appendix B.
Viscosity Repeatability: ±0.2%ofFullScaleRange