Brookfield DV2+Pro Viscometer Instrukcja Użytkownika Strona 19

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 101
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 18
Brookeld Engineering Labs., Inc. Page 19 Manual No. M03-165-F0612
II.6 Out of Range
The DV-II
+Pro gives indications for out-of-range operation. When % (Torque) readings exceed
100% (over-range), the display changes to that shown in Figure II-15; EEEE will also appear in
the display for shear stress:
cP EEEE 20.1C
10 RPM % EEEE
Figure II-15
Youmustchangeeitherspeedorspindletocorrectthiscondition.Ifyouoperateatspindlespeeds
that produce % (Torque)below10.0%,theDV-II+Proashesthe% (Torque), cP (Viscosity), SS
(Shear Stress) and SR(ShearRate)asshowninFigureII-16:
cP 12.4 20.1C
10 RPM % 8.2
Figure II-16
Negative%(Torque)willbedisplayedasshowninFigureII-17:
cP ---- 20.1C
10 RPM % -2.2
Figure II-17
Figure II-18 is an example of the printed output of each of the above conditions.
Normal Operation:
RPM=50 M=RV S=29 %=51.4 cP=10280 D/CM2=1285 1/SEC=12.3 T=20.1C Z00:30
Over-Range Operation (>100% torque) (see Fig. 15):
RPM=50 M=RV S=29 %=EEEE cP=EEEE D/CM2=EEEE 1/SEC=12.3 T=20.1C Z00:30
Under-Range Operation (<10% torque) (see Fig. 16):
? RPM=50 M=RV S=29 %=5.2 cP=1040 D/CM2=130 1/SEC=12.3 T=20.1C Z00:30
Negative Torque Operation (see Fig. 17):
RPM=50 M=RV S=29 %=-0.1 cP=---- D/CM2=---- 1/SEC=12.3 T=20.1C Z00:30
M = Torque Range T = TemperatureZ = Time
Figure II-18
Przeglądanie stron 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 100 101

Komentarze do niniejszej Instrukcji

Brak uwag