Brookfield DV-I Prime Instrukcja Użytkownika Strona 61

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 71
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 60
Brookeld Engineering Laboratories, Inc. Page 61 Manual No. M/07-022-D0613
Interpretation of Calibration Test Results:
When verifying the calibration of the DV-I PRIME, the instrument and viscosity standard uid
error must be combined to calculate the total allowable error.
The DV-I PRIME is accurate to (+/-) 1% of any Full Scale spindle/speed viscosity range.
Brookeld Viscosity Standards Fluids are accurate to (+/-) 1% of their stated value.
Calculate the acceptable range of viscosity using RVDV-I PRIME with
RV-3 Spindle at 2 RPM; Brookeld Standard Fluid 12,500 with a viscosity
of 12,257 cP at 25°C:
1) Determine Full Scale Viscosity Range using the AUTORANGE key on your instrument
or calculate with equation:
Full Scale Viscosity Range [cP] = TK
*
SMC
*
10,000
RPM
Where:
TK = 1.0 from Table D-2 (In Appendix D)
SMC = 10 from Table D-1 (In Appendix D)
Full Scale Viscosity Range
1
*
10
*
10,000
= 50,000 cP
2
The viscosity is accurate to (+/-) 500 cP (which is 1% of 50,000)
2) The viscosity standard uid is 12,257 cP. Its accuracy is (+/-) 1% of 12,257 or
(+/-) 122.57 cP.
3) Total allowable error is (122.57 + 500) cP = (+/-) 622.57 cP.
4) Therefore, any viscosity reading between 11,634.4 and 12,879.6 cP indicates that the
viscometer is operating correctly. Any reading outside these limits may indicate a viscometer
problem. Contact the Brookeld technical sales department or your local Brookeld dealer/
distributor with test results to determine the nature of the problem.
Calculate the acceptable range of viscosity using RVDV-I PRIME with
Small Sample adapter at 10 RPM; Brookeld Standard Fluid 12,500 with
a viscosity of 12,257 cP at 25°C:
1) Determine Full Scale Viscosity Range by pressing the AUTORANGE key on your
instrument. Display shows 25,000 cP. The viscosity is accurate to +/- 500 cP (which is
2% of 25,000 cP).
2) The viscosity standard uid is 12,257 cP
. Its accuracy is (+/-)1% of 12,257 or
(+/-) 122.57 cP.
3) Total allowable error is (122.57 + 500) cP = (+/-) 622.57 cP.
4) Therefore, any viscosity reading between 11,634.4 and 12,879.6 cP indicates that the
viscometer is operating correctly. Any reading outside these limits may indicate a viscometer
problem. Contact the Brookeld technical sales department or your local Brookeld dealer/
distributor with test results to determine the nature of the problem.
Example 2:
Example 1:
Przeglądanie stron 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 70 71

Komentarze do niniejszej Instrukcji

Brak uwag