Brookfield DV-I Prime Instrukcja Użytkownika Strona 22

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 71
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 21
Brookeld Engineering Laboratories, Inc. Page 22 Manual No. M/07-022-D0613
Using the UP and DOWN ARROW keys, the user enters a value for the minutes portion of
the time to stop. This value can be as high as 99 minutes.
2. When satised, the user presses the ENTER key again to enter the seconds setting display:
TIMED STOP
SET SEC’S:00
Figure II-18
Using the
UP and DOWN ARROW keys, the user enters a value for the seconds portion of the time
to stop. This value will be between 0 and 59 seconds.
The user presses the
ENTER key one more time at which point the viscometer will display the
following screen:
TIMED STOP:PRESS
MOTOR ON/OFF
Figure II-19
3. At this point, the user need only press the
MOTOR ON/OFF key to begin the timed stop
operation.
4. We will assume that the user pressed the MOTOR ON/OFF key to ON and is now presented
with the following screen for the duration of the timed run:
CP 123456789
MIN: 15 SEC: 13
Figure II-20
NOTE: When this mode has begun, a press of the MOTOR ON/OFF key will
interrupt the Timed Stop sequence and return the user to normal operation.
The seconds display will decrement from 59 to zero (0) in one (1) second intervals. When
seconds reaches zero (0), the minutes value will decrement by one (1) minute. This will
continue until all of the time has elapsed at which point the viscometer will display the
following screen:
cP 123456789
TIMED STOP DONE
Figure II-21
At this point, the viscometer will stop the motor and continue to display this screen until
the user presses the
UP or DOWN ARROW keys to view the Torque and Speed that were
current at the Timed Stop completion. This display would appear as follows:
Przeglądanie stron 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 70 71

Komentarze do niniejszej Instrukcji

Brak uwag