Brookeld Engineering Laboratories, Inc. Page 43 Manual No. M/07-022-D0613
A.2 Setup
1. Be sure that the Viscometer is securely mounted
to the Laboratory Stand, leveled and zeroed
with no cone or cup attached and 0% torque
is displayed.
2. Figure A-2 shows a typical water bath setup.
Connect the sample cup inlet/outlet ports to
the water bath inlet and outlet and set the
bath to the desired test temperature. Allow
sufcient time for the bath to reach the test
temperature. The temperature range of the
Sample Cup (CPE-44Y or CPE-44PY) is
0-100°C. Brookeld recommends a maximum
temperature of 80°C to allow for direct hand
contact for adjustment of the micrometer ring.
When using the sample cup at temperatures
near 0°C, be careful to avoid frost buildup on
the top surface of the cup; this could prevent a
proper t with the micrometer ring. Please refer
to the bath manual for the proper selection of
bath uid and tubing to ensure safe and proper
operation.
3. The Viscometer has been supplied with a special
cone spindle(s), which contains the Electronic
Gap Setting feature. The “CPE” part number
designation on the cone veries the Electronic
Gap Setting feature.
4. With the motor off, thread the cone spindle
by using the spindle wrench to secure the
viscometer coupling nut (see Figure A-3);
gently push up on the coupling nut and hold
this securely with the wrench. Thread the cone
spindle by hand. Note: Left Hand Threads.
5. Attach the cup, taking care not to hit the cone
with the cup (Figure A-4). There must be no
uid in the cup.
6. Option: The sample cup is available with an
optional purge tting. The user can connect
a dry gas line to this and put a blanket of dry
gas over the sample during measurement, if
desired.
Figure A-3
Spindle
Wrench
(CPE) Cone
These surfaces
must be clean!
Coupling Nut
Micrometer
Adjustment
Ring
Do Not hit the
CONE with the CUP!
Figure A-4
Sample
Cup
(CPE-44Y
or
CPE-44PY)
Bath/Circulator
Bath
Inlet
Bath
Outlet
Cup
Outlet
Cup
Inlet
Micrometer
Ring
Figure A-2