BrookeldEngineeringLabs.,Inc. Page26 ManualNo.M08-218-B0212
THERMOSTAT
TS: 20.0
T: 20.0
Status: 00
Permanent Indication of
Actual Temperature Value
Adjustable Temperature
Set Value
Error Indication
(00 = no error)
p / q
Info on Temperature
Value That Can Be Varied
The last adjusted set point TS is always saved when the rheometer is turned off. When the
rheometer is re-started, the Peltier will return to the last set point.
Modify the set point temperature with the
and
keys.
The temporary ashing of three points “• • •” conrms the acceptance of the new set tem-
perature value:
THERMOSTAT
TS: 45.0 ...
T: 20.0
Status: 00
p / q
Conrmation of Modied
Set Temperature Value
Adjustment of the actual temperature value starts automatically. The main menu can be
re-entered with the
and
keys.
VI.5.2 MAINMenu→ SingleMeasurement
This function measures shear stress or shear rate (CSS or CSR) at constant user dened
values. The user may select a choice of measurement method from the follwoing:
Shear Rate D [s
-1
]
Speed n [U/min] or [rpm]
Shear Stress Tau [Pa}
Torque
M[
‰] [1000‰ correspond to 50 mNm]
Before you can start the measurement itself, you have to input the following values:
• Selection of the measuring system used (see Section VI.3: Selecting from Lists).
• Selection of test method (see Section VI.3: Selecting from Lists). This is where you de-
cide if you want a controlled rate measurement (shear rate, RPM) or a controlled stress
measurement (shear stress, torque).
• Input of pre-set values (see Section VI.4: Input of Numerical Values and Alphanumer-
ic Texts).
- D [s
-1
] range of values depends on the measuring system
- n [rpm] 0.7 to 1,000 rpm
- Tau [Pa] range of values depends on the measuring system
- M[‰] 0-999 ‰
• Input of desired number of measuring points (see Section VI.4: Input of Numerical