V.2 FillingtheMeasuringSystems
Place the sample material on the bottom plate with the instrument head in the upper position.
Air bubbles in the sample will affect test results and should be avoided. Remove excess sample
from around the cone or plate with the provided spatula, taking care not to scratch the surface of
the bottom plate.
Remove excess sample from cone or plate edge.
Set the bath/circulator to the required temperature if measurements are to be taken under controlled
temperature. Do not begin the measurement until the temperature in the substance has stabalized at
set point, e.g. temperature measurement by Pt100 (See Section VI: Operation and Menu Systems).
Now, the measurement can begin.
Takecarewhenmeasuringathightemperature!Alwayscheckthetem-
peraturebeforeremovingtheconeorplatetoavoidburnedngers!
Besuretoholdthemeasuringsystemwithonehandwhenyoudisconnectthegeometry.
Droppingtheconeorplatecancausenickswhichwillaffectcalibration.
Clean the bottom measuring plate and measuring element carefully without using hard objects to
avoid scratches! Store the disassembled measuring element on a soft pad.
Alwaysdisconnecttheconeorplatebyliftingthecouplingcollarbefore
raisingtherheometerheadafterameasurement!