Brookeld Engineering Labs., Inc. Page 23 Manual No. M98-350-J0912
VISCOMETER TORQUE CONSTANT MODEL CODE
MODEL TK ON DV-E SCREEN
LVDV-E 0.09373 LV
RVDV-E 1 RV
HADV-E 2 HA
HBDV-E 8 HB
SPECIAL ORDER TORQUE SPRINGS
VISCOMETER TORQUE CONSTANT MODEL CODE
MODEL TK ON DV-E SCREEN
2.5xLVDV-E 0.2343 2.5LV
5xLVDV-E 0.4686 5LV
1/4 RVDV-E 0.25 1/4RV
1/2 RVDV-E 0.5 1/2RV
2xHADV-E 4 2HA
2.5xHADV-E 5 2.5HA
2xHBDV-E 16 2HB
2.5xHBDV-E 20 2.5HB
Table C-2
This equation can be used to calculate the maximum viscosity that can be measured when using a
specic speed/spindle combination.
Full Scale Viscosity Range (FSR) = TK * SMC *
10,000
RPM
The units for FSR are in centipoise (cP). An example is shown in Appendix D.