QTS MATERIALS EVALUATION SYSTEM OPERATOR MANUAL
Page 14
Select utility : < .. Monitor load .. >
b) < ..
Cal. distance
.. >
Select utility: Cal. distance .. >
Press 'ENTER' Set calibrated probe
just above platen
Screw in 80mm calibration probe and use the up and down
↑↓
keys to position
probe approximately 3mm above platen.
Upon selection of this option the 'CALIBRATE' LED on QTS25 straining frame will
be lit
Press 'ENTER' Nulling Cell
Beam Creeping
When the probe touches the platen the beam stops.
Entered Calibrated Length
Key in 80 and press 'ENTER'
New Distance
Calibration Set
The beam moves upwards to allow the calibration probe to be removed and the
display reverts to :
QTS Controller READY
"Time" "Day" "Date"
c) < ..
Exit from menu
.. >
Select utility : < .. Exit from menu
Press 'ENTER' QTS Controller READY
"Time" "Day" "Date"
d)
Statistics Functions
.. >
Select utility: < Statistics Functions >
Press 'ENTER'
Select option:
Invalidate Result
Recover Result
Erase All Statistics
Exit from Menu