Appendix B - Troubleshooting
If no power to the instrument:
Check:
• Userreplaceablefuses:two(2)fuses,4amp,5x20mm,time-lag
Fuses located within power entry module on the rear of the instrument.
If fuses blow continually, serious damage to the instrument could result. Contact repair
services immediately.
Probe won’t attach:
Check thread on probe and probe shaft - remove any dirt
Base table won’t move:
Ensure locking knob is loose
• Ensureneadjustnutisloose
Base table will not travel to the lowest position:
• Loosenneadjustnutandrotatebasetable
Instrument is unsteady:
Ensure that bench top is stable
• AdjusttheLFRAfeet
Probe doesn’t move when start is pressed:
Ensure emergency stop button is not pressed in (rotate clockwise)
Ensure test mode is not Static Load
Ensure probe is not in contact with sample (or base table)
Display shows “REMOTE OPERATION”
This will always show when the computer interface cable is connected to the LFRA. To
use the LFRA without the computer program, the serial cable must be disconnected.