Brookfield DV1 Viscometer Instrukcja Użytkownika Strona 44

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 63
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 43
Brookeld Engineering Laboratories, Inc. Page 44 Manual No. M14-023
VISCOMETER MODEL
TORQUE
CONSTANT TK
MODEL CODE ON
DV1 SCREEN
DV1MLV 0.09373 LV
2.5xDV1MLV 0.2343 2.5LV
5xDV1MLV 0.4686 5LV
1/4 DV1MRV 0.25 1/4RV
1/2 DV1MRV 0.5 1/2RV
DV1MRV 1 RV
DV1MHA 2 HA
2xDV1MHA 4 2HA
2.5xDV1MHA 5 2.5HA
DV1MHB 8 HB
2xDV1MHB 16 2HB
2.5xDV1HB 20 2.5HB
Table D-2
e full scale viscosity range for any DV1 model and spindle may be calculated using the equation:
Full Scale Viscosity Range [cP] = TK
*
SMC
*
10,000
RPM
where:
TK = DV1 Torque Constant from Table D-2
SMC = Spindle Multiplier Constant from Table D-1
e Shear Rate calculation is:
Shear Rate (
1
/
Sec
) = SRC
*
RPM
where:
SRC = Shear Rate Constant from Table D-1