2. Spindle jammed below zero
Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for
repair
Display Reading Over 100
1. Overrange “EEE” (in %, cP, and SS)
Change spindle and/or speed
Viscometer Will Not Return to Zero
1. Viscometer is not level
Check with spindle out of the sample
Adjust the laboratory stand
2. Pivot point or jewel bearing faulty
Perform calibration check
Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for
repair
Display Reading Will Not Stabilize
1. Check for erratic spindle rotation
Verify power supply
Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for
repair
2. Bent spindle or spindle coupling
Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for
repair
3. Temperature fluctuation in sample fluid
Use temperature bath control
4. Special characteristic of sample fluid
Refer to Appendix C