Brookeld Engineering Labs., Inc. Page 1 Manual No. M02-313-G0812SPECIALISTS IN THE MEASUREMENT ANDCONTROL OF VISCOSITY TEL508-946-6200FAX508-946-6
Brookeld Engineering Labs., Inc. Page 10 Manual No. M02-313-G0812I.6 Safety Symbols and Precautions Safety Symbols The following explains saf
Brookeld Engineering Labs., Inc. Page 11 Manual No. M02-313-G0812 HOLD TIME This key selects time entry mode (wait time before spindle rotates).S
Brookeld Engineering Labs., Inc. Page 12 Manual No. M02-313-G0812 II. GETTING STARTEDII.1 Power On Turn the power ON using the switch located
Brookeld Engineering Labs., Inc. Page 13 Manual No. M02-313-G0812II.2 Cone Spindle Selection and Setting Raise the viscometer handle to its high
Brookeld Engineering Labs., Inc. Page 14 Manual No. M02-313-G0812Spindle 04CALIBRATE? YES/NOFigure II-4If you are using the CAP Viscometer for th
Brookeld Engineering Labs., Inc. Page 15 Manual No. M02-313-G0812II.4 Temperature Control Setting Press the TEMP key and the current temperature
Brookeld Engineering Labs., Inc. Page 16 Manual No. M02-313-G0812Note: With a run time of zero, the hold time will not be usedThe time required fo
Brookeld Engineering Labs., Inc. Page 17 Manual No. M02-313-G08122. Pressing and holding the STOP/ESCAPE key during power up selects the special f
Brookeld Engineering Labs., Inc. Page 18 Manual No. M02-313-G0812 III. OPERATION The CAP 2000+ Viscometer rotates a precisely machined cone spin
Brookeld Engineering Labs., Inc. Page 19 Manual No. M02-313-G0812Table III-2: Full Scale Range Viscosity for HIGH TORQUE CAP 2000+ ViscometerCone
Brookeld Engineering Labs., Inc. Page 20 Manual No. M02-313-G0812Table III-3: Accuracy for LOW TORQUE CAP 2000+ ViscometerCone Number ≤ 500 RPM10-
Brookeld Engineering Labs., Inc. Page 21 Manual No. M02-313-G0812III.3 Calibration Verication Accuracy of the CAP 2000+ Viscometer can be veri
Brookeld Engineering Labs., Inc. Page 22 Manual No. M02-313-G0812 Figure III-1a Figure III-1b Figure III-1c Figure III-1d 4. Run
Brookeld Engineering Labs., Inc. Page 23 Manual No. M02-313-G0812Table III-5: Viscosity Standard Fluids for Calibration of CAP Spindles on LOW TOR
Brookeld Engineering Labs., Inc. Page 24 Manual No. M02-313-G0812Note: A cone calibration should be performed when: 1) using a new cone for the
Brookeld Engineering Labs., Inc. Page 25 Manual No. M02-313-G0812Note: The calibration process may be cancelled at any time prior to pressing RUN
Brookeld Engineering Labs., Inc. Page 26 Manual No. M02-313-G0812III.6 Making Viscosity MeasurementsThe following procedure is recommended for m
Brookeld Engineering Labs., Inc. Page 27 Manual No. M02-313-G0812Notes: a. Lower the handle gently. DO NOT FORCE THE CONE ONTO THE PLATE.
Brookeld Engineering Labs., Inc. Page 28 Manual No. M02-313-G0812When advanced sample analysis is required, Brookeld CAPCALC application software
Brookeld Engineering Labs., Inc. Page 29 Manual No. M02-313-G0812 APPENDIX A - Variables in Viscosity MeasurementsAs with any instrument measuremen
Table of Contents I. INTRODUCTION ...
Brookeld Engineering Labs., Inc. Page 30 Manual No. M02-313-G0812 Viscometer Related Variables• Most uid viscosities are found to be non-Newtonia
Brookeld Engineering Labs., Inc. Page 31 Manual No. M02-313-G0812 APPENDIX B - CommunicationsPrinter Output - CAP 2000+The cable connection on the
Brookeld Engineering Labs., Inc. Page 32 Manual No. M02-313-G0812CAP 2000+ TRANSMIT/RECEIVE COMMANDS FOR COMPUTER COMMUNICATIONCOMMAND RECEIVEDCAP
Brookeld Engineering Labs., Inc. Page 33 Manual No. M02-313-G0812Note: All multiplication and division operations performed on any of the command
Brookeld Engineering Labs., Inc. Page 34 Manual No. M02-313-G08121 Bit 2 is set in response to an R command (request for information) only.2 If an
Brookeld Engineering Labs., Inc. Page 35 Manual No. M02-313-G0812 APPENDIX C - Online Help and Additional Resourceswww.brookeldengineering.com**
Brookeld Engineering Labs., Inc. Page 36 Manual No. M02-313-G0812 APPENDIX D - Packing Instructions to Return a Brookeld CAP Viscometer for
Brookeld Engineering Labs., Inc. Page 37 Manual No. M02-313-G0812 APPENDIX E - Warranty Repair and ServiceWarrantyBrookeld Viscometers are guara
Brookeld Engineering Labs., Inc. Page 5 Manual No. M02-313-G0812 I. INTRODUCTIONThe CAP 2000+ Series Viscometers are medium to high shear rate in
Brookeld Engineering Labs., Inc. Page 6 Manual No. M02-313-G0812Please check to be sure that you have received all components and that there is no
Brookeld Engineering Labs., Inc. Page 7 Manual No. M02-313-G0812I.3 Specications Torque Range: Low 797-7,970 dyne•cm (designated on serial
Brookeld Engineering Labs., Inc. Page 8 Manual No. M02-313-G0812I.4 Dimensional Details
Brookeld Engineering Labs., Inc. Page 9 Manual No. M02-313-G0812I.5 InstallationDO NOT lift the viscometer by the handle or head! LIFT only by