Brookfield Engineering Labs., Inc. page 16 Manual No. M14-223
II.4.5.1 Assembly of FTK Temperature Control Device (Water Jacket)
This section describes assembly and connection of the following measuring systems:
Figure II-2
Components:
1 Threaded joint for mounting flange 4 Hose conncetion for thermal oil discharge
2 Thermostatting chamber 5 Pt100 temperature sensor
3 Threaded joint for sample cup 6 Hose connction for thermal oil inflow
Optional temperature control devices available:
Liquid temperature control device FTK for use in cylinder measuring systems in the
temperature range of -10°C ... +90°C (temperature control by liquid thermostat).
If the optional KE cooling device is used, the temperature control device FTK can be
operated in the temperature range of -20°C ... +180°C.
Caution!
Temperature control device
The temperature control device FTK
without KE cooling unit must not be used
beyond the temperature range of -10°C ...
+90°C!
Figure II-3